Hit and False-Alarm Rates of Selected ABR Indices in Differentiating Cochlear Disorders From Acoustic Tumors We performed a retrospective study of 26 patients with acoustic tumors and 26 patients with otologically diagnosed cochlear pathology to determine the sensitivity (hit rate), specificity (false-alarm rate), and efficiency of six auditory brainstem response indices. In addition, a utility value was determined for each of these six indices. The ... Research Article
Research Article  |   March 01, 1996
Hit and False-Alarm Rates of Selected ABR Indices in Differentiating Cochlear Disorders From Acoustic Tumors
 
Author Affiliations & Notes
  • Frank E. Musiek, PhD
    Section of Otolaryngology, Dartmouth-Hitchcock Medical Center, One Medical Center Drive, Lebanon, NH 03756-0001
  • Cynthia A. McCormick
    University of Tennessee, Knoxville
  • Raymond M. Hurley
    Louisiana State University Medical Center, New Orleans
Article Information
Hearing Disorders / Balance & Balance Disorders / Research Articles
Research Article   |   March 01, 1996
Hit and False-Alarm Rates of Selected ABR Indices in Differentiating Cochlear Disorders From Acoustic Tumors
American Journal of Audiology, March 1996, Vol. 5, 90-96. doi:10.1044/1059-0889.0501.90
History: Received June 30, 1994 , Accepted October 28, 1994
 
American Journal of Audiology, March 1996, Vol. 5, 90-96. doi:10.1044/1059-0889.0501.90
History: Received June 30, 1994; Accepted October 28, 1994

We performed a retrospective study of 26 patients with acoustic tumors and 26 patients with otologically diagnosed cochlear pathology to determine the sensitivity (hit rate), specificity (false-alarm rate), and efficiency of six auditory brainstem response indices. In addition, a utility value was determined for each of these six indices. The I–V interwave interval, the interaural latency difference, and the absolute latency of wave V provided the highest hit rates, the best A’ values and good utility. The V/I amplitude ratio index provided high specificity but low sensitivity scores. In regard to sensitivity and specificity, using the combination of two indices provided little overall improvement over the best one-index measures.