Threshold Measurement for Patients With Tinnitus
Author Notes
Article Information
Hearing Disorders
Article   |   March 1997
Threshold Measurement for Patients With Tinnitus
American Journal of Audiology, March 1997, Vol. 6, 52-56. doi:10.1044/1059-0889.0601.52
History: Received September 7, 1995 , Accepted April 1, 1996
American Journal of Audiology, March 1997, Vol. 6, 52-56. doi:10.1044/1059-0889.0601.52
History: Received September 7, 1995; Accepted April 1, 1996

This study compared the effectiveness of pulsed tones (200 ms on/200 ms off) and continuous tones (1 to 2 s) for measuring thresholds in patients with tinnitus. Sixty veterans with tonal tinnitus and a history consistent with noise-induced hearing loss participated. Half were assigned to a pulsed-tone group; half were assigned to a continuous-tone group. Air-conduction thresholds were measured twice for each patient in a single session for octave intervals between 250 Hz and 4000 Hz. Results indicated no significant difference between groups for test-retest differences, but overall more presentations were required to measure thresholds with continuous tones than with pulsed tones at 4000 Hz. Moreover, the continuous-tone group had, on average, more false-positive responses than the pulsed-tone group. This study supports the clinical belief that pulsed tones are preferred over continuous tones for measuring thresholds for persons with tinnitus.